Author: Hwang, J.G.
Paper Title Page
MOPMR008 Development of Beam Position Monitor for a Heavy-ion Linac of KHIMA 238
 
  • J.G. Hwang
    KIRAMS/KHIMA, Seoul, Republic of Korea
  • G. Hahn, T.K. Yang
    KIRAMS, Seoul, Republic of Korea
 
  Funding: This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIP) (no. NRF-2014M2C3A1029534).
The carbon and proton beams are produced by the electron cyclotron resonance ion source with the energy of 8 keV/u and it is accelerated up to 7 MeV/u by the RFQ and IH-DTL. The accelerated beam is injected on the synchrotron through the medium energy beam transport (MEBT). In the MEBT line of KHIMA, the stripline beam position monitor (BPM) is installed to measure the beam trajectory and orbit jitter before the beam injection at the synchrotron. It is also used to measure the phase information such as a bunch length for the de-buncher tuning in MEBT line. The BPM has the position resolution of 100 um with the diameter of 40 mm. The design study is performed and it is fabricated. In order to confirm the performance of the beam position monitor, the measurement of position accuracy and calibration by using wire test-bench, and the beam test with proton beam from MC-50 in KIRAMS are performed.
 
DOI • reference for this paper ※ DOI:10.18429/JACoW-IPAC2016-MOPMR008  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)  
 
MOPMR009 Development of Large Aperture Faraday-Cup for LEBT of KHIMA. 241
 
  • J.G. Hwang
    KIRAMS/KHIMA, Seoul, Republic of Korea
  • T.K. Yang
    KIRAMS, Seoul, Republic of Korea
 
  Funding: This work was supported by the National Research Foundation of Korea(NRF) grant funded by the Korea government(MSIP) (no. NRF-2014M2C3A1029534).
Since an aperture of a low energy beam transport line of the KHIMA is quite large, 100 mm, to minimize an uncontrolled beam loss, large aperture Faraday-cup with the diameter of 100 mm is installed to measure the beam current from the electron cyclotron resonance ion source (ECR-IS) and to identify the ion species using analyzing magnet. The suppression ring is designed to reduce the repelling electrons for an accurate measurement. The Faraday-cup has the cooling channel with the heat capability of 100 W to recover the heat from the ion beam for safety during the operation. In order to reduce the noise propagation from the cooling channel, the cooling channel is insulated with the cup. In this presentation, we show the physical modeling, mechanical aspect for design the large aperture Faraday-cup, and the result of in-beam test with the ECR-IS in KHIMA.
 
DOI • reference for this paper ※ DOI:10.18429/JACoW-IPAC2016-MOPMR009  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)  
 
MOPMR010 The Development of Scintillating Screen Detector for Beam Monitoring at the KHIMA Project 244
 
  • S.Y. Noh, S.D. Chang, J.G. Hwang
    KIRAMS/KHIMA, Seoul, Republic of Korea
  • G. Hahn, T.K. Yang
    KIRAMS, Seoul, Republic of Korea
 
  Funding: NRF-2014M2C3A1029534
It is important to measure the beam propeties such as position, size and intensity, when we control the medical beam qualities, So we developed the scintillation screen monitor used for beam profile monitoring and it will be installed at High Energy Beam Transport(HEBT) section to measure the beam parameters. This system consists of a terbium-doped gadolinium oxysulfide(Gd2O2S:Tb) phosphor screen and high speed charge coupled device camera. The CCD camera has the maximum 90 frame rate and 659 X 494 pixel resolution. This Camera is mounted at distance of 260mm from the center of the scintillation screen and with the angle of 45 degree to the scintillation screen which is mounted at the angle of 45 degree to the beam axis. The image analysis program was written in National Instruments LabVIEW using IMAQ driver. To reduce the image processing time, we optimized the prcessing flow and used LabVIEW built-in function. To evaluate this system, we measured the beam size and center position of the beam at KIRAMS on 50MeV cyclotron. In this paper, we present the manufacture of beam profile system based on a scintillating screen monitor and the in-beam test results of it.
 
DOI • reference for this paper ※ DOI:10.18429/JACoW-IPAC2016-MOPMR010  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)  
 
FRXAA01 Korea Heavy Ion Medical Accelerator Project 4243
 
  • G.B. Kim, G. Hahn, W.T. Hwang, H. Yim
    KIRAMS, Seoul, Republic of Korea
  • J.G. Hwang, C.H. Kim, C.W. Park
    KIRAMS/KHIMA, Seoul, Republic of Korea
 
  The Korea Heavy Ion Medical Accelerator (KHIMA) project is to develop 430-MeV/u heavy ion accelerator and therapy systems for medical applications. The accelerator system includes ECRIS, injector linac, synchrotron, beam transport lines, and treatment systems. The accelerator system is expected to provide stable beams very reliably, and there should be special cares and strategies in the machine construction and operations. This presentation covers all issues mentioned above.  
slides icon Slides FRXAA01 [10.869 MB]  
DOI • reference for this paper ※ DOI:10.18429/JACoW-IPAC2016-FRXAA01  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)