TUPC095 Beam Diagnostics for Commissioning the HEBT and Gantry Sections of the HIT Medical Accelerator diagnostics, ion, extraction, controls 1281
 
  • M. Schwickert, A. Reiter
    GSI, Darmstadt
  The HIT medical accelerator at Heidelberg, Germany, is the first dedicated heavy-ion cancer therapy facility in Europe, consisting of a two-stage injector Linac followed by a compact synchrotron. It features three treatment places: two horizontal beam lines, where treatment will be carried out from 2008 using proton and carbon beams, and the first 360° rotating heavy-ion Gantry structure. The accelerator sections of this facility were designed and constructed by GSI, which thereafter was in charge of the commissioning. By now, the required medical beam quality has been achieved in both horizontal beam lines, and beam commissioning of the Gantry structure has started. In this contribution we describe the technical layout of beam diagnostic devices and present measurement data taken in high-energy beam transport lines and patient treatment places.