Funding: Pole Mecatech/Biowin/SPW RW - Convention 8150: CardiAmmonia In 2020, IBA has started the design, construction, tests and industrialization of a new proton cyclotron for the low energy range, the Cyclone® KEY, for PET isotope production (18F, 13N, 11C) for neurology, cardiology or oncology imaging. It is a compact and fully automated isochronous cyclotron accelerating H− up to 9,2 MeV. Based on the successful design history and return of experience of the Cyclone® KIUBE, the Cyclone® KEY design has been focused on compactness (self-shielding enabled), cost effectiveness and ease of installation, operation and maintenance. The innovative design consists in the interoperability of the different subsystems: the magnet, the RF system, the vacuum system, the ion source, the stripping extraction, and target changers (with up to three targets). First beam tests results will also be presented. |
||
![]() |
Slides WEAO03 [2.848 MB] | |
DOI • | reference for this paper ※ doi:10.18429/JACoW-CYCLOTRONS2022-WEAO03 | |
About • | Received ※ 22 December 2022 — Revised ※ 11 January 2023 — Accepted ※ 01 February 2023 — Issue date ※ 11 April 2023 | |