Paper |
Title |
Page |
MOPP020 |
First Tests Using Sipm Based Beam Loss Monitors at the European XFEL |
127 |
|
- T. Wamsat, P.A. Smirnov
DESY, Hamburg, Germany
|
|
|
The European XFEL MTCA based Beam Loss Monitor System (BLM) is composed of about 450 monitors using photomultiplier tubes (PMTs). BLMs installed in the SASE undulator intersections show high signals at electron energy higher 16 GeV or photon energy higher 14 keV due to background synchrotron radiation which directly affects the PMT. The amplitude of this signal can get that high that, also without using any scintillating material, the BLMs get blind for real losses. Also different lead arrangements did not shield the signal sufficiently. First tests show that a Silicon photomultiplier (SiPM) is not affected. Also there are several advantages to use SiPM, they are cheaper by factor of 40 and operating voltage is below 45V. First test will be presented and how it can get implemented in the existing BLMs and BLM system.
|
|
DOI • |
reference for this paper
※ https://doi.org/10.18429/JACoW-IBIC2019-MOPP020
|
|
About • |
paper received ※ 04 September 2019 paper accepted ※ 08 September 2019 issue date ※ 10 November 2019 |
|