Pulse Scanning System of the Medical Beam and the System of Recording the Position of the Beam and the Distribution of the Irradiation Dose of the Object |
217 |
|
- I.P. Yudin
JINR/VBLHEP, Dubna, Moscow region, Russia
- S.I. Tyutyunnikov
JINR, Dubna, Moscow Region, Russia
|
|
|
This work describes the pulse scanning system of the medical beam and the system of recording the position of the beam and the distribution of the irradiation dose of the object. Our report deals with the construction of the carbon beam transport line for biomedical research at the NICA accelerator complex, JINR, Dubna. We are discussed the compilation and realization of the plan of treating a tumor located at a depth up to 30 cm. The hardware realization of the beam scanning is shown. We describe the wire chamber and read out electronics which were developed for on-line monitoring of the "intensive" extracted beams of relativistic nuclei. The system has been tested in several acceleration runs with deuteron beams with the intensity up to 1010 1/s and carbon ion beams. The system allows one to make multiple measurements of the two-dimensional beam intensity distribution and beam position during the beam spill.
|
|