Paper | Title | Page |
---|---|---|
WEPPP050 | Quick scanning Channel-Cut crystal monochromator for millisecond time resolution EXAFS at HEPS | 229 |
|
||
The design and capabilities of a Quick scanning Channel-Cut monochromator (QCCM) for HEPS are presented. The quick scan and step scan are realized by a torque motor directly driven Bragg axis, controlled by a servo controller. This design allows easy and remote control of the oscillation frequency and angular range, providing comprehensive control of QXAFS measure-ments. The cryogenically cooled Si (311) and Si(111) crystals, which extends the energy range from 4.8 keV-45 keV. The dynamic analysis verifies the rationality of the mechanical structure design. The device was fabri-cated and tested, results show an oscillation frequency up to 50Hz with a range of 0.8°, and a resolution of 0.2 arcsecond in step scan mode. This device demonstrates the feasibility of large range quick scan and step scan by a single servo control system.
Quick scanning Channel-Cut crystal monochromator |
||
DOI • | reference for this paper ※ doi:10.18429/JACoW-MEDSI2023-WEPPP050 | |
About • | Received ※ 02 November 2023 — Revised ※ 05 November 2023 — Accepted ※ 08 November 2023 — Issued ※ 08 January 2024 | |
Cite • | reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml) | |