THPOPT065 Operation of X-Ray Beam Position Monitors with Zero Bias Voltage at Alba Front Ends 2747
 
  • J. Marcos, U. Iriso, V. Massana, R. Monge, D. Yépez
    ALBA-CELLS Synchrotron, Cerdanyola del Vallès, Spain
 
  Blade-type X-ray Beam Position Monitors (XBPMs) are customarily operated with a negative bias voltage applied to the blades in order to prevent the transference of photoelectrons between the blades, and hence to maximize the signal at each blade and to avoid cross-talk. This was the selected approach at ALBA since the start of its operation for users in 2012. However, over the years the insulation provided by the ceramic pieces separating the blades from the support structure has degraded progressively, giving rise to an ever-increasing leakage current not related with the photon beam to be monitored. On 2020 the level of these leak currents had already become comparable to the photocurrents generated by the photon beam itself, making the readings from many of the XBPMs unreliable. Following the example from other facilities, we decided to remove the bias voltage from the blades and to test the performance of the XBPMs under these conditions, with such good results that we apply this method also for the new, non degraded, XBPMs. In this paper we present the approach used at ALBA to analyse XBPM data, and our experience operating them with zero bias voltage.  