Author: Liu, J.
Paper Title Page
MOPAB121 Progress Towards Soft X-Ray Beam Position Monitor Development 438
 
  • B. Podobedov, C. Eng, S. Hulbert, C. Mazzoli
    BNL, Upton, New York, USA
  • D. Donetski, K. Kucharczyk, J. Liu, R. Lutchman
    Stony Brook University, Stony Brook, New York, USA
 
  Funding: Work supported by Brookhaven Science Associates, LLC under Contract No. DE-SC0012704 with the U.S. Department of Energy.
X-ray beam position monitors (BPMs) are instrumental for storage ring light sources, where they reliably provide positional measurements of high-power beams in hard X-ray beamlines. However, despite a growing need, coming especially from coherent soft X-ray beamlines, non-invasive soft X-ray BPMs have not been demonstrated yet. We are presently working on a funded R&D proposal to develop a non-invasive soft X-ray BPM with micron-scale resolution for high-power white beams. In our approach, multi-pixel GaAs detector arrays are placed into the beam halo and beam position is inferred from the pixel photocurrent levels. Presently, the first detector array prototypes have been manufactured and are being prepared for low-power beam tests. The mechanical design of a BPM test-stand, which will be installed in the 23-ID canted soft X-ray undulator beamline at NSLS-II, is well under way. In addition, we are developing new algorithms of beam position calculation which take full advantage of extended multi-pixel detector arrays. In this paper we will review our design choices and discuss recent progress.
 
DOI • reference for this paper ※ https://doi.org/10.18429/JACoW-IPAC2021-MOPAB121  
About • paper received ※ 03 June 2021       paper accepted ※ 13 July 2021       issue date ※ 28 August 2021  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)