MOPAB181 | Non-Delivery Time Reduction at MedAustron | 613 |
|
||
Funding: Funding by the NÖ WIRTSCHAFTS- UND TOURISMUSFONDS under grant number WST3-F-5033232/001-2020. MedAustron is a cancer treatment center in Austria providing proton and carbon ion beams to three clinical and one non-clinical research beam lines. The slow extraction of particles from the synchrotron follows a third order resonance extraction scheme. Currently, for every change of extraction energy a new spill needs to be generated. Besides the beam-on time of the particle delivery, every spill is also comprised of non-delivery time components e.g. the multiturn injection, acceleration or magnet conditioning. For small tumor target volumes, this non-delivery time is the major contribution to the overall treatment time. A dedicated performance improvement project (supported with a grant from the state of lower Austria) was executed with the goal to reduce these non-delivery times without affecting important clinical beam parameters such as the beam size or penetration depth. The implemented reduction of the non-delivery time >50% could be achieved, resulting in beam-on time reductions for reference treatment plans between 25% (largest proton PTV) and 58% (smallest carbon PTV). Results of commissioning efforts, technical details and the achieved optimizations will be presented. |
||