Author: Guo, H.D.
Paper Title Page
THPRB118 Study on the Influence of the Range Shifter Material in a Scanning Nozzle for Proton Therapy Based on Monte Carlo Method 4100
 
  • Y.C. Yu, H.D. Guo, Y.Y. Hu, X.Y. Li, Y.J. Lin, P. Tan, X.D. Tu, L.G. Zhang
    HUST, Wuhan, People’s Republic of China
 
  Range shifter plays a key role in decreasing the energy of the proton beam to realize shallow tumours treatment with the scanning nozzle in Huazhong University of Science and Technology Proton Therapy Facility (HUST-PTF). To control the transverse scattering and decrease the damage to healthy tissue caused by secondary particle, influence of the range shifter material was studied. In this paper, the Monte Carlo software Geant4 and FLUKA are applied to analyse the transport process of proton beam in the range shifters made of six different materials: PMMA, Lexan, Lucite, Polyethylene, Polystyrene, and Wax. The beam spot sizes at the iso-center with or without range shifter was calculated for the HUST-PTF scanning nozzle. The relationship between the thickness of the range shifters of the six materials and the proton energy was obtained. The secondary neutron yield at the end of the nozzle was also analysed.  
DOI • reference for this paper ※ https://doi.org/10.18429/JACoW-IPAC2019-THPRB118  
About • paper received ※ 14 May 2019       paper accepted ※ 23 May 2019       issue date ※ 21 June 2019  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)