Short period in-vacuum, permanent magnet undulators operating at cryogenic temperatures are being developed worldwide to serve as brilliant and coherent light sources for medium energy storage rings. A hybrid cryogenic permanent magnet undulator (CU) with PrFeB magnets has now been designed and constructed at NSRRC [1]. To characterize the performance and to determine magnetic field errors after cool down poses some technical chal-lenges compared to room temperature undulators. A new system combining a Hall probe and a stretched wire has been designed to measure the field integrals, trajectory, phase errors, and K value under low temperature and vacuum conditions. Field measurements in this cryogenic undulator will be performed around 77 K as well as at room temperature, making temperature dependent calibra-tion of the Hall probes necessary. The main features and improvement of the measurement and calibration system are presented.
|
|