JACoW logo

Joint Accelerator Conferences Website

The Joint Accelerator Conferences Website (JACoW) is an international collaboration that publishes the proceedings of accelerator conferences held around the world.


RIS citation export for THPVA140: Superconducting Gantry Design for Proton Tomography

TY - CONF
AU - Oponowicz, E.
AU - Owen, H.L.
ED - Schaa, Volker RW
ED - Arduini, Gianluigi
ED - Pranke, Juliana
ED - Seidel, Mike
ED - Lindroos, Mats
TI - Superconducting Gantry Design for Proton Tomography
J2 - Proc. of IPAC2017, Copenhagen, Denmark, 14–19 May, 2017
C1 - Copenhagen, Denmark
T2 - International Particle Accelerator Conference
T3 - 8
LA - english
AB - Precise proton therapy planning can be assisted by augmenting conventional medical imaging techniques with proton computed tomography (pCT). For adults this requires an incident proton energy up to around 330 MeV, requiring superconducting magnets if an imaging gantry is to replace a conventional 230-250 MeV gantry in the same space. Here we present optics considerations for a superconducting gantry to deliver 330 MeV protons within the context of the future Christie Hospital proton therapy centre, where it is proposed to increase the proton energy in the future with a booster linac.
PB - JACoW
CP - Geneva, Switzerland
SP - 4795
EP - 4797
KW - proton
KW - dipole
KW - quadrupole
KW - linac
KW - optics
DA - 2017/05
PY - 2017
SN - 978-3-95450-182-3
DO - 10.18429/JACoW-IPAC2017-THPVA140
UR - http://jacow.org/ipac2017/papers/thpva140.pdf