Author: De Arcangelis, D.
Paper Title Page
MOPMW007 On the Calibration Measurement of Stripline Beam Position Monitor for the ELI-NP Facility 411
 
  • D. De Arcangelis, F. Cardelli, A. Mostacci, L. Palumbo
    University of Rome La Sapienza, Rome, Italy
 
  Stripline Beam Position Monitor (BPM) will be installed in the Compton Gamma Source in construction at the ELI Nuclear Physics facility in Romania. A test bench for the calibration of BPM has been built to characterize the device with stretched wire measurement in order to get the BPM response map. A full S-parameters characterisation is performed as well to measure the electrical offset with the "Lambertson method". This paper discusses the extensive simulations performed with full 3D electromagnetic CAD codes of the above measurements to investigate measurement accuracy, possible measurement artefacts and the beam position reconstruction.  
DOI • reference for this paper ※ DOI:10.18429/JACoW-IPAC2016-MOPMW007  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)