Author: Benveniste, E.
Paper Title Page
TUPC129 A Beam Position System for Hadrontherapy Facilities 1323
 
  • A. Faus-Golfe, C. Belver-Aguilar, C. Blanch Gutierrez, J.J. García-Garrigós
    IFIC, Valencia, Spain
  • E. Benveniste, M. Haguenauer, P. Poilleux
    LLR, Palaiseau, France
 
  Funding: MICINN-FPA:AIC10-D-000518
Essential parts of the needed instrumentation for the beam control in the Hadrontherapy accelerators are the Beam Position Monitors (BPM). The measurement of the beam position in Hadronterapy accelerators become more important at the secondary transport lines towards the patient room where this parameter must be completely determined. The BPM described in this paper is a new type of BPM based on four scintillating fibers coupled to four photodiodes to detect the light produced by the fibers when intercepting the beam. We present here the study of the different photodiodes able to read the light emitted by the scintillating fiber, the tests performed in order to find the most suitable photodiode to measure the beam position from the variations in the beam current, the mechanical design and the corresponding acquisition electronics.