Author: Wei, K.
Paper Title Page
MOPA04 The Beam Instruments for HIMM@IMP 33
 
  • T.C. Zhao, Y.C. Chen, J.M. Dong, Y.C. Feng, X.C. Kang, M. Li, S. Li, W.L. Li, W.N. Ma, R.S. Mao, H.H. Song, K. Song, Y. Wang, K. Wei, Z.G. Xu, Y. Yan, Y. Yin, Z.L. Zhao
    IMP/CAS, Lanzhou, People’s Republic of China
 
  HIMM(Heavy Ion Medical Machine)is a synchrotron based accelerator for cancer therapy in Wuwei city, China. It is composed of 2 ion sources, LEBT, cyclotron, MEBT, a synchrotron, HEBT and therapy terminals. The commissioning of HIMM is completed .At present, electrical safety, electromagnetic compatibility and performance testing of medical devices have been passed, and now enters the clinical tests phase. The beam diagnositics(BD) devices for HIMM are designed and produced by IMP BD department .An overview of the integrated devices is presented, and the common beam parameters in the different parts of the accelerator facility are reviewed including intensity measurement, beam profile, emmitance, energy and so on with the related detectors such as the View Screen, Faraday Cup, Radial Detector, Multi-wires, Phase Probe, Wire Scanner, DCCT, ICT, BPM, Schottky, Slit, Beam Stopper, Beam Halo Monitor, Multi-channel Ionization Chamber. Additionally, the RF-KO for beam extraction, the strip foil with automatic control system as well as the detectors for terminal therapy are described.  
DOI • reference for this paper ※ https://doi.org/10.18429/JACoW-IBIC2018-MOPA04  
About • paper received ※ 05 September 2018       paper accepted ※ 13 September 2018       issue date ※ 29 January 2019  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)