Paper | Title | Page |
---|---|---|
WEP079 | Effect of Heat Load on Cryo-Cooled Monochromators at the European X-Ray Free-Electron Laser: Simulations and First Experimental Observations | 502 |
|
||
European XFEL (EuXFEL) generates high-intensity ultra-short pulses at MHz repetition rate. At hard X-ray instruments, cryo-cooled silicon monochromators are used to reduce pulse bandwidth. Here, first experimental observations during commissioning of a cryo-cooled monochromator at Materials Imaging and Dynamics (MID) instrument are presented and compared with heat flow simulations. A thermal relaxation time is estimated and compared with arrival time interval between pulses. This provides the repetition rate tolerable for stable operation of monochromator. | ||
DOI • | reference for this paper ※ https://doi.org/10.18429/JACoW-FEL2019-WEP079 | |
About • | paper received ※ 19 August 2019 paper accepted ※ 25 August 2019 issue date ※ 05 November 2019 | |
Export • | reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml) | |