The electron beam asymmetry observed at the Photo Injector Test Facility at DESY in Zeuthen (PITZ) was traced back to multipole kicks in the gun section, namely around the location of the coaxial power coupler and the main solenoid. Several dedicated studies have been performed to quantify the kick location and strength. Based on these studies, two designs of correction quadrupole coils were proposed. The coils were fabricated and tested with an electron beam. The second updated design implies a two quadrupole setup on a frame installed around the gun coaxial coupler close to the main solenoid centre location. Skew and normal quadrupole magnets are powered independently, enabling flexibility in electron beam manipulations. By means of this setup, a more symmetric beam was obtained at several screens. This led also to more equal measured horizontal and vertical phase spaces and to even smaller overall emittance values. Some details of the gun quadrupole designs, magnetic measurements, and results of electron beam measurements including emittance optimization will be reported.
|
|