A   B   C   D   E   F   G   H   I   J   K   L   M   N   O   P   Q   R   S   T   U   V   W   X   Y   Z    

Merinov, I.G.

Paper Title Page
TUPCH072 New Generation Streak Camera Design and Investigation 1175
 
  • A.M. Tron
    LPI, Moscow
  • T.V. Gorlov, I.G. Merinov
    MEPhI, Moscow
 
  The only method for electron bunch duration monitoring with a resolution in the order of 10 fs and less is the method of photochronography of the bunch incoherent radiation in the frequency range, for example, of visible light and at realizing streak camera with new principles of its operation*. In the paper the streak camera design for measuring both the electron bunches and x-ray pulses duration with the mentioned temporal resolution is presented. The results of the camera investigation, with photoelectron dynamics simulation taking into account space-charge effect and impact of the surface roughness of a spherical photocathode of the 20-50 micrometers radius (forming a modulating gap of spherical configuration) on the camera resolution, are presented and discussed.

*A. M. Tron and I. G. Merinov. Method of bunch radiation photochronography with 10 femtosecond and less resolution. http://www.physics.ucla.edu/PAHBEB2005/talks/10oct2005/wg2/atron.pdf