  Pohang Accelerator Laboratory (PAL) built a PAL-ITF at the end of 2012 to successfully complete PAL-XFEL in 2015. The PAL-ITF is equipped with various kinds of diagnostic equipment to produce high-quality electron bunches. An ICT and a Turbo-ICT were installed in the PAL-ITF. A Faraday Cup is installed at the end of the linear accelerator. These days, the quantity of electric charge occasionally is measured using a BPM Sum value. This paper focuses on the processes and results of electric charge quantity measurements using ICT, Turbo-ICT, FC and BPM. The PAL-ITF is equipped with Stripline-BPM. It is important to find a way to minimize measurement errors that can appear in the process of installing or measuring the BPM. For this, PAL-ITF separately measured the BPM electrode sensitivity and minimized BPM measurement errors through generally calibrating BPM devices by applying Lambertson's Method. A plan was made to minimize BPM measurement errors through applying the BPM electrical calibration method for BPM devices to be used by the PAL-XFEL. This paper examines the processes for checking the performance of the S-BPM installed in the PAL-ITF and the results of its measurements.  