TUPRO094 Magnetic Field Evaluation of Multipole Permanent Magnets by Harmonic Coil with Novel Calibration Technique 1259
SUSPSNE097   use link to see paper's listing under its alternate paper code  
 
  • R. Kitahara, Y. Fuwa, Y. Iwashita
    Kyoto ICR, Uji, Kyoto, Japan
 
  Quadrupole magnets for ILC final focus should be strong enough with the restriction on the external radius to let the disrupted out-going beam pass by, while vibration of the magnetic center has to be highly avoided to keep the nm sized beam focusing stable at the interaction point a few meter downstream from the lens. Gluckstern's 5-ring PMQ singlet seems a good candidate for this point of view. In order to fabricate a good 5-ring singlet, property of each ring has to be good enough. A harmonic coil system, which has 24-bit ADC’s for high resolution, was developed. Current noise level of the system is less than 10-5, which is supposed to be improved by reducing mechanical vibration of the ball bearings. We demonstrated the evaluation method of coil wire position with magnetic field from pin point magnet, so that the accuracy of the method was comparable to um scale. We measured the prototype 5-ring PMQ singlet and evaluated harmonic components. This result was compared with the data measured at KEK.  