Author: Grossmann, M.
Paper Title Page
THPHA098 Development of a PXI Based Test Stand for Automatization of the Quality Assurance of the Patient Safety System in a Proton Therapy Centre 1604
 
  • P. Fernandez Carmona, M. Eichin, M. Grossmann, F. Heimann, H.A. Regele, D.C. Weber, R. van der Meer
    PSI, Villigen PSI, Switzerland
 
  At the Centre for Proton Therapy at the Paul Scherrer Institute a cyclotron, two gantries and a fixed beamline are being used to treat tumours. In order to prevent non-optimal beam delivery, an interlock patient safety system (PaSS) was implemented that interrupts the treatment if any sub-system reports an error. To ensure correct treatment, the PaSS needs to be thoroughly tested as part of the regular quality assurance as well as after each change. This typically required weeks of work, extensive beam-time and may not comprehensively detect all possible failure modes. With the opportunity of the installation of a new gantry, an automated PaSS test stand was developed that can emulate the rest of the facility. It consists of a NI PXI chassis with virtually unlimited IOs synchronously stimulated or sampled at 1MHz, a set of adapters to connect each type of interfaced signal and a runtime environment. We have also developed a VHDL based formal language to describe stimuli, assertions and specific measurements. We present the use of our test stand in the verification and validation of the PaSS, showing how its full quality assurance, including report generation was reduced to minutes.  
poster icon Poster THPHA098 [1.561 MB]  
DOI • reference for this paper ※ https://doi.org/10.18429/JACoW-ICALEPCS2017-THPHA098  
Export • reference for this paper using ※ BibTeX, ※ LaTeX, ※ Text/Word, ※ RIS, ※ EndNote (xml)