MOPA24 Photon Beam Position Monitor at SIAM Photon Source photon, undulator, insertion, electron 104
 
  • P. Sudmuang, S. Boonsuya, S. Chaichuay, P. Klysubun, S. Krainara, H. Nakajima, S. Rugmai, N. Sumano, N. Suradet
    SLRI, Nakhon Ratchasima, Thailand
 
  Photon beam position monitors (PBPM) have been designed and installed in the beamline front-ends at Siam Photon Source (SPS). Up till now, these blade-type PBPMs have been successfully installed at three bending magnet and an insertion device (planar undulator) beamlines. Its performance has been tested and compared with that of the electron beam position monitor. The achieved resolution is found to be better than 3 μm. The obtained PBPM data proved to be extremely invaluable in the investigation of the sources of the observed beam positional fluctuation, and for compensation of the orbit perturbation caused by undulator gap change. In this paper, the details of the calibration procedure will be presented. Various factors affecting reading of the signal such as back scattering effect, choice of bias voltage, and temperature variation have been investigated and the results will be discussed herewith.