TY - CPAPER AU - Stulle, F. AU - Bergoz, J.F. TI - Turbo-ICT Pico-Coulomb Calibration to Percent-level Accuracy J2 - Proc. of FEL2015 AB - We report on the calibration methods implemented for the Turbo-ICT/BCM-RF. They allow to achieve percent-level accuracy for charge and current measurements. Starting from the Turbo-ICT/BCM-RF working principle, we discuss scientific fundaments of calibration and their practical implementation in a test bench. Limits, both principle and practical, are reviewed. Achievable accuracy is estimated. PB - JACoW CY - Geneva, Switzerland SP - 118 EP - 121 KW - resonance KW - impedance KW - laser KW - plasma KW - network DA - 2015/12 PY - 2015 SN - 978-3-95450-134-2 DO - 10.18429/JACoW-FEL2015-MOP041 UR - http://accelconf.web.cern.ch/AccelConf/FEL2015/papers/mop041.pdf ER -