JACoW logo

Joint Accelerator Conferences Website

The Joint Accelerator Conferences Website (JACoW) is an international collaboration that publishes the proceedings of accelerator conferences held around the world.


RIS citation export for WEPP010: Design and Simulation of a Cavity BPM for HUST Proton Therapy Facility

TY  - CONF
AU  - Li, J.Q.
AU  - Chen, Q.S.
AU  - Fan, K.
AU  - Tang, K.
AU  - Tian, P.
ED  - Schaa, Volker RW
ED  - Jansson, Andreas
ED  - Shea, Thomas
ED  - Olander, Johan
TI  - Design and Simulation of a Cavity BPM for HUST Proton Therapy Facility
J2  - Proc. of IBIC2019, Malmö, Sweden, 08-12 September 2019
CY  - Malmö, Sweden
T2  - International Beam Instrumentation Conferenc
T3  - 8
LA  - english
AB  - In proton therapy facility, non-destructive beam diagnostic devices are essential for on-line measurement during the patient treatment. To meet the clinical requirement, the beam current becomes ultra-low of the order of nano-ampere, which is a great challenge to non-destructive beam diagnostics because of the extremely low signal level. Compared with conventional non-destructive beam diagnostic devices, the cavity beam position monitor (BPM) has a high shunt impedance to get enough power levels, so a cavity BPM system is designed for HUST-PTF. It is made up of two resonant cavities called reference cavity and position cavity, respectively. Both cavities are simulated and optimized by CST Microwave Studio and Particle Studio. Finally, the electronics of cavity BPM we plan to use is shown.
PB  - JACoW Publishing
CP  - Geneva, Switzerland
SP  - 530
EP  - 533
KW  - cavity
KW  - proton
KW  - coupling
KW  - simulation
KW  - impedance
DA  - 2019/11
PY  - 2019
SN  - 2673-5350
SN  - 978-3-95450-204-2
DO  - doi:10.18429/JACoW-IBIC2019-WEPP010
UR  - http://jacow.org/ibic2019/papers/wepp010.pdf
ER  -